DIGITAL FIBER SPECTROMETER (USB3000-ILX511)
Order Part No.:#2066

|
USB3000 is applicable for spectral detection within the wavelength range between 200nm and 1100nm. Because of the high stability and high performance, these portable instruments can greatly satisfy the increasing need for scientific research, industrial manufacture, online detections and so on.
USB3000 is capable of measuring
Spectral power distribution, Absorbance, transmission, reflection, Relative irradiance, Chromaticity coordinates, Color temperature, Color rendering index, Color tolerance, Color difference, Color purity, Main wavelength, Luminous flux, Luminance, Irradiance power and Luminous effciency |
 |
.etc of various kinds of samples, light sources and display devices. Besides they can also be utilized to achieve photochromic analysis of
Florescence lamps, High pressure gas discharge light sources, Halogen-tungsten lamps, medical light sources, Semi-conductor lighting devices and Display devices as CRT, LCD, PDP, ELD, VFD.
USB3000 can offer the optimal solutions for spectral detections. With different system configurations and sampling accessories, USB3000 can achieve high spectral resolution up to 0.2nm, stable wide, narrow and line spectrums with different optical interfaces. Furthermore, with the easy-to-use basic operating software and high effcient SDK, end users could easily achieve various kinds of spectral dectection systems to meet their own goals. |
|
|
Mechanical parameters |
|
Dimensions (mm) |
95*68.5*36 |
|
Weight (g) |
210 |
|
Detector specification |
|
Detector |
Sony ILX511 2048 Linear CCD array |
|
Detector wavelength response range (nm) |
200-1100 |
|
Pixels |
2048 |
|
Pixel size (µm) |
14*200 |
|
Pixel well depth (number electrons) |
105 |
|
SNR |
300:1 (full signal) |
|
A/D resolution |
12bit |
|
Dark noise (RMS counts) |
50 |
|
Corrected linearity |
99.8% |
|
Sensitivity (photons/count) |
400nm : 130; 600nm : 60 |
|
Optical bench |
|
Optical system |
f/4, asymmetrical crossed Czerny-Turner |
|
Focal length (mm) |
45 (input); 70 (output) |
|
Entrance aperture (µm) |
5, 10, 15, 20, 50 .etc width slits or fiber (customizable) |
|
Grating |
various kinds of scoring grating or holographic grating with different grating constants and flare wavelengths (customizable) |
|
Cylindrical lens |
optional |
|
Optical filter |
bandpass or longpass filters to eliminate secondary spectrum |
|
Fiber and fiber connection |
NA 0.22, SMA905 (customizable) |
|
Spectroscopic |
|
Wavelength range |
depend on optical grating |
|
Optical resolution (nm) |
0.2-10 FWHM (depend on grating constant, width of slit and detector model) |
|
SNR |
300:1 (full signal) |
|
Wavelength tolerance (nm) |
about 0.2 |
|
Dark noise (RMS counts) |
50 |
|
Integration time |
10µs-64s |
|
Dynamic range |
2.0*108(system); 1300:1 (single acquisition) |
|
Stray light |
<0.05% at 550nm |
|
Electronics |
|
Power consumption |
250 mA/5 V DC |
|
Data transfer speed |
full spectrum into memory every 5 ms at USB 2.0 (16ms at USB 1.1) |
|
Computer |
|
Operating system |
Windows 2003/XP/Vista |
|
Interface |
USB 2.0 (480Mbps), compatible with USB 1.1 |
|
Pre-defined system components |
|
Name |
Customizable parts |
Remark |
|
Slit |
Width (µm) |
10, 20, 50, 100 .etc |
|
Grating |
Lines per unit (L/mm) |
300, 600, 1200, 1800 |
|
Working wavelength range(nm) |
200-1100 (User defined, about 650 range with 600 lines/mm grating) |
|
Fiber |
Core diameter (µm) |
9, 50, 200, 400, 600 .etc |
|
Working wavelength range(nm) |
200-1100 |
|
Outer protection layer |
Normal or enhanced jacketing |
|
Interface type |
SMA 905 (standard) or FC, ST .etc |
|
Detector |
Model |
Sony ILX511 |
|
Cylindrical lens |
Optional |
Enhance system¡¯s light gathering ability |
|
Optional components |
|
Name |
Remark |
|
Light source |
Used for wide range spectral measurement or¡¡wavelength calibration for example halogen tungsten, xenon or argon light sources |
|
Integration sphere |
Used to acquire uniformed light |
|
Reflection measurement accessory |
Components used for reflection spectrum measurements for example standard reflection board or specialized clamp |
|
Transmission measurement accessory |
Components used for transmission spectrum measurement for example sample holder and various kinds of cuvettes |
|
Packing list |
|
Name |
Amount |
|
Digital Fber Spectrometer |
USB3000-ILX511 |
|
Fiber with SMA 905 interface |
1 |
|
Mini USB data transfer wire |
1 |
|
CD of Software and SDK |
1 |
System Drawing

Fig 1. System drawing
Detection Result
Fig 2. Detection result of standard low- voltage mercury lamp( 1200L/mm grating and 50um slit)
Fig 3. Detection result of standard low- voltage mercury lamp( 1200L/mm grating and 9um slit)
Fig 4. Detection result of HG-1 mercury argon calibration light source ( 600L/mm grating and 50um slit)
Left:USB6000's Right:A world famous brand's

|